How Do I Change Stimulation Intensity?
Stimulation intensity is expressed as a percentage of the patient's Motor Threshold (MT). Stimware automatically calculates the treatment intensity based on the selected protocol and the patient's saved MT value. Authorized users can adjust the stimulation intensity when clinically appropriate.
Change Stimulation Intensity
- Load the patient profile.
- Open the treatment session.
- Review the patient's Motor Threshold value.
- Locate the Intensity setting.
- Increase or decrease the intensity as required.
- Confirm the updated treatment settings before beginning or resuming stimulation.
How Is Intensity Calculated?
Treatment intensity is typically prescribed as a percentage of the patient's Motor Threshold.
For example:
- Motor Threshold (MT): 50%
- Treatment Intensity: 120% MT
Stimware automatically calculates the corresponding stimulation output based on these values.
Before Adjusting Intensity
Always verify:
- The correct patient is loaded
- The patient's Motor Threshold is up to date
- The selected treatment protocol is correct
- Coil positioning is accurate
Important Considerations
- Changes to stimulation intensity should follow the prescribed treatment plan.
- Intensity adjustments may affect treatment delivery and patient comfort.
- Always review treatment settings before starting stimulation.